Nude Ionization Vacuum Gauge Tubes
• Ion (Hot Filament) Gauge Tube
• Dual Filaments
• Cage Grid
• EB Degas Compatible
UHV-24 Nude Bayard-Alpert Type Ion Gauge
Varian Vacuum (Agilent) Equivalent
Granville Phillips (MKS) Equivalent
Operating Data
|
UHV-24 Series
|
Torr Maximum Pressure:
|
1x10-3
|
Torr X-Ray Limit:
|
2x10-11
|
Sensitivity +/-20%:
|
25/Torr
|
Emission:
|
4 ma.
|
Filament Voltage:
|
3-4.5 Volts
|
Filament Current:
|
2-4 amps
|
Filament Potential:
|
+30 Volts
|
Collector Potential:
|
0 Volts
|
Grid Potential:
|
+180 Volts
|
Degass, Watts:
|
30-40 by EB
|
Bakeout
Temperature:
|
450°C |
|
Contact LDS Vacuum with any questions:
Phone (407)862-4643 via Fax (407)862-8723 or
info@ldsvacuum.com
LDS Vacuum Products is a 100% woman owned company and certified Woman Owned Small Business (WOSB) The logos and/or trademarks appearing on this page are not owned by or registered to LDS but, rather, designate products sold by LDS, and are owned and/or registered to the manufacturers or suppliers of the products.